Awarded
CAN: Procurement of Repl SO2 NDIR Gas Analysers
Descriptions
The National Physical Laboratory (NPL) requirement is for replacement NDIR and paramagnetic sensor analysers, with the following capabilities: 1.Capable of measuring the following gases below their respective amount fractions: a.SO2 - 1 % b.O2 - 25 % c.CO - 5 % d.CO2 - 10 % 2.The analyser shall be CE/UKCA certificated. 3.There must be a datasheet, manual and user instructions 4.Analog output (4-20 mA or 0-10 Vdc) is required to allow connectivity to exiting automatic analysers and LabView integration. 5.Flowrate: ≤ 1 L min-1 6.Linearity error: ≤ 1.5 % of span 7.Repeatability: ≤ 1 % of span 8.The analyser must be suitable for 19" rack installation 9.Permitted pressure: 800 - 1200 kPa 10.Detection limit (4 σ): ≤ 1 % of span
Timeline
Published Date :
Deadline :
Tender Awarded :
Awarded date :