Awarded
Four Optically Pumped Magnetometers
Descriptions
Cardiff University require a set of four optically pumped magnetometers for performing magnetoencephalography recordings directly from scalp. Each of them should be no larger than 13x17x25 mm, have a field sensitivity of less than 15 fT/√Hz in the frequency band between 5 and 100 Hz, and less than 25 fT/√Hz in simultaneous dual axis operation. They should also have a tolerance to background magnetic fields of 200nT. Additionally, the kit should include an adapter board (NI-9205 or similar).