Open Closing in a 1 month
Ion Chromatography Systems to replace the current Ion Chromatograph
Descriptions
The Environmental and Industrial Hazards (EIH) Laboratory at the Institute of Naval Medicine requires two Ion Chromatography Systems to replace the current Ion Chromatograph which has reached the end of its useful life and to provide additional capability according to new requirements.
The systems will primarily be used for analysis of anions in raw, potable and bottled water samples in accordance with the requirements of service doctrine1, 2, 3 and UK drinking water legislation4, 5, but other matrices will include environmental samples such as groundwater, surface water and wastewater.
One Ion Chromatograph must be capable of analysing 9 anions (fluoride, chlorite, chloride, nitrite, chlorate, bromide, nitrate, sulphate and phosphate) using conductivity detection.
The other Ion Chromatograph must be capable of analysing bromate in samples with elevated levels of chloride with post column derivatisation and absorbance detection. Specifically, post column addition of potassium iodide followed by detection of triiodide using a uv-vis detector.
The systems must meet the requirements for trueness, precision and limit of detection detailed in The Water Supply (Water Quality) Regulations 2018.
See Invitation to Tender for further details.
Timeline
Published Date :
Deadline :
Contract Start :
Contract End :
Tender Regions
CPV Codes
38432200 - Chromatographs
Keywords
chromatographs
chromatography instruments
GC units
HPLC analyzers
chromatographic systems
liquid chromatography devices
column chromatography equipment
Tender Lot Details
2 Tender Lots
Workflows
Status :
Procedure :
Suitable for SME :
Nationwide :
Assign to :
Tender Progress :
Details
Notice Type :
Tender Identifier :
Tenderbase ID :
Low Value :
High Value :
Buyer Information
Name :
Procurement contact
Name :
Phone :
Email :
Possible Competitors
1 Possible Competitors