Awarded

Cryomechanical test frame for materials qualification

Descriptions

Envisaged scope of supply

• Dry cryostat compatible with digital image correlation, conventional contact-based extensometery and point sensors (e.g., strain gauges). Some compromise in performance is permissible when conducting measurements using DIC setups requiring windows. Target temperature is as low as possible, minimum temperature requirement is 10 K (with DIC).
• Temperature control to within +- 1 K.
• Modularity and flexibility to support a range of diagnostics and potential future modifications.
• Sufficiently large enough to accommodate standard sized testing in line with ASTM E8 and E606
• Must also be compatible with small scale materials testing (e.g., SSJ3 sample).
• Robust high vacuum chamber with appropriate thermal radiation shielding.
• Must be able to support a fully aligned rigid load string.
• The load string should be flexible and allow the testing of complex geometry samples and subcomponents.
• Diagnostics to help optimise the performance of the system. I.e. be flexible enough to optimise conduction pathways.
• Quick sample changeover is highly desirable. Preferred testing throughput of at least 2 tests a day for tensile testing.
• Must be able to support long term testing programmes in excess of 100 tests.
• Robust and repeatable temperature control system and sensor interface with sample.
• Must interface with a four columns Zwick SS-CF Kappa 100 kN load frame or similar.
• Pull bar feed throughs with bellows.
• Vacuum pumping solutions via suitable flanges.
• Support a minimum vacuum of 1 x 10-6 mbar.
• Ability to backfill vacuum using dry gas (e.g. from compressed N2 cylinder)
• Windows and ports to facilitate and permit digital image correlation (DIC).
• Sufficient access to permit DIC calibration.
• Doors and/or sufficient access to permit sample exchange.
• Be able to support uniform temperature testing.
• Be able to actively monitor temperature at specific locations on the sample.
• Be able to apply a thermal gradient across the specimen with 50 K difference from either end of the gauge length of the specimen (Stretch target).
• Feedthroughs for in-chamber services (e.g., cryo pump, diagnostics, etc… ).
• Easy to use control system.
• Control system allows vacuum and cooling to automatically start via timer (to optimise test throughput).
• Extensometer feed throughs and supports.
• Heating solution mechanical supports (if required), access, and service feedthroughs.
• Internal heating to speed up warm up.
• Be able to operate for prolonged periods of time in order to support fatigue testing.
• Cryogenic TMF testing capability is also desirable but not essential.

Timeline

Published Date :

5th Feb 2026 7 days ago

Deadline :

N/A

Contract Start :

N/A

Contract End :

N/A

Keywords

on-site assembly

equipment alignment

mechanical anchor setup

control panel setup

hardware commissioning

fixture installation

non-software integration

structural bracket installation

piping system installation

lighting fixture mounting

Tender Lot Details & Award Criteria

1 Tender Lots

Let’s Get you Started ✍

Get to see all tender details more briefly

Already have an account ?

Workflows

Status :

Awarded

Procedure :

N/A

Suitable for SME :

N/A

Nationwide :

No

Assign to :

Tender Progress :

0%

Details

Notice Type :

Tender

Tender Identifier :

IT-378-246-T: 2024 - 001

Tenderbase ID :

310724019

Low Value :

£100K

High Value :

£1000K

Buyer Information

Address :

Liverpool Merseyside , Merseyside , L13 0BQ

Website :

N/A

Procurement Contact

Name :

Tina Smith

Phone :

0151 252 3243

Email :

tina.smith@shared-ed.ac.uk