Lightricity is looking for a close-match solar simulator that provides a compact, bench top unit complete with IV measurement. The solar simulator will be used for benchmarking and quality control. Required specifications (AAA or better) Spectral: Class AM0 / AM1.5 spectrum Method of measurement: Demonstrate spectral match with calibrated spectroradiometer (range 300nm to 1800nm); A or better Uniformity ±2% over target diameter nominal 60mm (standard IEC 904-9); A or better Method of measurement: Scan as per accepted standard Temporal stability: ≤±1%; A or better Method of measurement: Demonstrate stability by acquiring IV curves and showing Isc reproducibility of ≤±1%. Specimen holder: > 5 cm x 5 cm; Should include test probes (<0.5mm diameter) and temperature controller (-20 °C- 100°C) The supplier will be selected based on value for money (70% weighting), warranty period (15% weighting) and customer support (15% weighting). Apply by email via the email address in the About the Buyer section of this notice.